In response to Jim Phillips' comments about filter specifications, it is true
that longpass, shortpass, and bandpass filters are specified with an
orientation orthogonal to the light source, but dichroics are normally
specified at the wavelength they split at when mounted at 45 degrees, the
normal situation with dichroics.

Mike Fox
Colorado State University


Home Page Table of Contents Sponsors Web Sites
CD ROM Vol 2 was produced by staff at the Purdue University Cytometry Laboratories and distributed free of charge as an educational service to the cytometry community. If you have any comments please direct them to Dr. J. Paul Robinson, Professor & Director, PUCL, Purdue University, West Lafayette, IN 47907. Phone:(317) 494-0757; FAX (317) 494-0517; Web http://www.cyto.purdue.edu EMAIL robinson@flowcyt.cyto.purdue.edu