Re: Sorting Question

Daryl Webb (dwebb@waite.adelaide.edu.au)
Fri, 20 Jun 1997 14:54:38 +0930 (CST)

Vincent Falco laments:
>
>
> My problem is this:During some sorts I get ,what I will call a spurious extra
> drop. To explain more,there are times I sort to polycarb filters.When all is
> well I get a single spot which contains the desired cell population.When all
> is supposed not well I get my desired single spot with cells and just about a
> quarter of inch from the desired spot I observe a spurious spot,which does
> not contain the desired cells.

Hi, I've gottta have a stab at this one... Its a few years since I ran a
sorter in my pevious life, but this sounds to me like a case of uncharged
"satellite drops" merging with one of your charged drops. Someone will
tell me if I'm off on a tangent :-)

I'm guessing 3 drop sorts here. When the drops breakoff the joining neck
forms a small "satellite droplet" , which depending upon phasing and
frequency tends to merge either with the cell above or below it.

(this is from deep memory, but there is a good discussion in Flow Cyt and
cell sorting 2nd ed,... The Yellow one... by Myron Melamid (sp) and others)

You can see all of this through the observation microscope, have a look
to see which cell the satellite merges with when all is well, and I'll
bet it goes the other way when you see the "doublet". I think you have
an uncharged droplet merging with a charged cell, reducing the effective
charge, causing less deflection and the appearance of the second spot.

I think its a pressure/frequency thing, but check in the old yellow bible
as its well covered.

Best of luck

-- 

Daryl Webb (dwebb@waite.adelaide.edu.au) Dept. of Plant Science, Waite Institute University of Adelaide, Glen Osmond S.A. 5064 Australia. Voice:61_8 8303 7426 Fax:61_8 8303 7102


Home Page Table of Contents Sponsors E-Mail Archive Web Sites

CD-ROM Vol 3 was produced by Monica M. Shively and other staff at the Purdue University Cytometry Laboratories and distributed free of charge as an educational service to the cytometry community. If you have any comments please direct them to Dr. J. Paul Robinson, Professor & Director, PUCL, Purdue University, West Lafayette, IN 47907. Phone: (765)-494-0757; FAX(765) 494-0517; Web http://www.cyto.purdue.edu , EMAIL cdrom3@flowcyt.cyto.purdue.edu