> Hello,
>
> I have been doing some E.Coli analysis and sorting and was
> wondering if anyone had a favorite sheath fluid to use for these
> bugs. I am looking to make them as happy as possible during sort
> cloning. Thanks for any tips.
Many of the commercial preps are unsuitable for bacteria because of
the high sodium content - they are designed to keep animal (blood)
cells happy. I haven't done sorting with coli but Micrococcus luteus
seem ok (i.e. grow afterwards) when sorted in a sheath fluid
containing 150mM KCl and 10mM Hepes (pH 6.8 with KOH). This was with
cells that weren't exactly happy to begin with :
Kaprelyants AS, Mukamolova GV, Davey HM, and Kell DB: Quantitative
analysis of the physiological heterogeneity within starved cultures of
Micrococcus luteus using flow cytometry and cell sorting. Appl. Env.
Microbiol. 62:1311-1316, 1996.
Haze
----------------------------------------------------------
| Hazel Marie Davey hlr@aber.ac.uk |
|Sefydliad y Gwyddorau Biolegol*Inst. Biological Sciences|
|Prifysgol Cymru * University of Wales|
| ABERYSTWYTH, Ceredigion, CYMRU / WALES SY23 3DA |
| http://pcfcij.dbs.aber.ac.uk/index.htm |
CD-ROM Vol 3 was produced by Monica M. Shively and other staff at the
Purdue University Cytometry Laboratories and distributed free of charge
as an educational service to the cytometry community.
If you have any comments please direct them to
Dr. J. Paul Robinson, Professor & Director,
PUCL, Purdue University, West Lafayette, IN 47907.
Phone: (765)-494-0757;
FAX(765) 494-0517;
Web